Best view resolution 1024x768
ICP-OES
Model: Thermo ScientificTM iCAPTM 7600
Model: Thermo ScientificTM iCAPTM 7600
(ICP-OES) Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES) is an established and powerful technique for simultaneous analysis and quantification of trace elements in both liquid and solid samples. It is based on the core advanced technologies to solve the challenges of the most demanding laboratory applications. It is a type of emission spectroscopy that uses the inductively coupled plasma to produce excited atoms and ions that emit electromagnetic radiation at wavelengths characteristic of a particular element. During analysis the intensity of the light emitted at specific wavelengths is measures and used to determine the concentration of the element(s) of interest. The thermal excitation sources can populate a large number of different energy levels for several different elements at the same time where all of the excited atoms and ions can then emit their characteristic radiation at the same time. This results in the flexibility to choose from several different emissions concurrently and allows detection of multiple elements simultaneously.
Applications:
ICP can be used in the quantitative analysis in the following areas such as Physical sciences and Nano-composites, environmental pollution control studies, Forensics, Industrials metals, natural materials such as rocks, minerals, soil, sediment, air, water, plant and animal tissue; pure and applied geochemistry, mineralogy, agriculture, forestry, animal husbandry, chemical ecology, environmental sciences and food industry, medical sciences etc The analysis of elements not readily identified by AAS such as Boron, Phosphorus, Titanium, and Zirconium etc can be done with ICP.
Specifications:
Applications:
ICP can be used in the quantitative analysis in the following areas such as Physical sciences and Nano-composites, environmental pollution control studies, Forensics, Industrials metals, natural materials such as rocks, minerals, soil, sediment, air, water, plant and animal tissue; pure and applied geochemistry, mineralogy, agriculture, forestry, animal husbandry, chemical ecology, environmental sciences and food industry, medical sciences etc The analysis of elements not readily identified by AAS such as Boron, Phosphorus, Titanium, and Zirconium etc can be done with ICP.
Specifications:
Type | iCAP 7600 ICP-OES Duo |
Wavelength Range | 166 to 847nm |
Detector Type | High performance solid-state CID86 chip |
Auxiliary Gas | Mass flow control, 0-2 L/min |
Data Acquisition Modes | Precision mode, Speed mode, Sprint mode |
Depth (Metric) | 750mm |
Full-frame Capability | Yes |
Height (Metric) | 590 mm |
Nebulizer Gas | Mass flow control, 0-2 L/min |
Peristaltic Pump | 4-channel, mini peristaltic pump, Speed: 0 to 125rpm |
Plasma Gas | Mass flow control, 0-2 L/min |
Plasma Viewing | Duo or Radial |
RF Source | 27.12MHz solid state, 750 to 1600w output power (Duo restricted to 1350w) |
Spectral Bandpass | 7pm at 200nm |
Spectrometer | Simultaneous echelle type, 52.91 grooves/mm ruled grating, 383mm effective length, 9.5° UV fused silica cross dispersion prism |
Width (Metric) | 840mm |
Item Description | Standard: Sprint valve aqueous (4mL loop), Optional: Sprint valve organic (3mL loop) |
Standard Sampling Kit | Concentric glass nebulizer, Glass cyclonic spray chamber, Semi-demountable EMT torch, 1.5mm bore quartz center tube (Radial), 2mm bore quartz center tube (Duo). |
Features | Sprint Valve, Intelligent introduction and rinse, Sprint data acquisition mode, Get ready: automatic performance checks, Gas flow (plasma, auxillary and nebulizer), Plasma power, Pump speed, Viewing height (Radial) |
(NB: The detection limit and sensitivity varies from element to element)